Aust Endod J ; 49(3): 592-598, 2023 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-37592880

RESUMEN

The aim of this study was to compare the cyclic fatigue resistance of nickel-titanium instruments inside single and double-curved canals at different temperatures. 160 HyFlex EDM #20.05 (HEDM), VDW.ROTATE #20.05 and #25.06, Mtwo #25.06 were randomised (n = 10) for the dynamic cyclic fatigue tests according to the curvature (i.e. single and double) at 20° ± 1°C and 35° ± 1°C. The number of cycles to fracture (NCF) was analysed by two-way ANOVA with p < 0.05. Fatigue resistance of all instruments significantly decreased at body temperature in single and double curvatures, except for HEDM in double curvature. The NCF was significantly lower in double curvature than single at both temperatures for all files, except for VDW.ROTATE #20.05 at 35° ± 1°C. Within the study limitations, temperature significantly impaired cyclic fatigue resistance of all files except HEDM #20.05 in double curvature. Similarly, double curvature had a detrimental effect on cyclic fatigue resistance of all files except for VDW.ROTATE #20.05 at body temperature.


Asunto(s)
Instrumentos Dentales , Níquel , Titanio , Temperatura Corporal , Aleaciones Dentales , Diseño de Equipo , Falla de Equipo , Ensayo de Materiales , Preparación del Conducto Radicular , Resistencia Flexional
2.
Clin Oral Investig ; 27(4): 1793-1798, 2023 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-36454355

RESUMEN

OBJECTIVES: To evaluate the influence of different heat treatments and temperatures on the cyclic fatigue resistance of 2Shape instruments (Micro Mega, Besancon, France) with the same design. MATERIALS AND METHODS: A total of 80 2Shape TS1 (#25/.04) and 80 TS2 (#25/.06) files with different heat treatments (no heat-treated, NHT; C-Wire; T-Wire; CM-Wire) were tested at room (25° ± 1 °C) and body (37° ± 1 °C) temperatures in 16-mm stainless steel artificial canal with a curvature of 60° and 5 mm of radius. Files were tested in continuous rotation at 300 rpm using a customized device. Cyclic fatigue resistance was expressed in the number of cyclic to fracture (NCF). The phase transformation temperature of each file was determined by differential scanning calorimetry. Data were statistically analyzed using the two-way ANOVA and the Bonferroni post-hoc test with p < 0.05. RESULTS: All CM-Wire TS1 and TS2 files showed higher NCF than the other groups at both temperatures (p < 0.001). C-Wire TS2 showed higher resistance than NHT and T-Wire TS2 (p < 0.001), with no significant differences between the last two. Body temperature significantly decreased NCF of all tested files (p < 0.05) except for NHT and T-Wire TS1. CONCLUSIONS: Body temperature negatively affected the cyclic fatigue resistance of C-Wire and CM-Wire TS1 and of all TS2 files compared with room temperature. The CM-Wire instruments exhibited the highest cyclic fatigue resistance of all tested files. CLINICAL RELEVANCE: Heat-treated nickel-titanium files can differently perform according to environmental temperature and file dimensions. The CM-Wire 2Shape prototypes displayed the highest flexural resistance in all experimental conditions.


Asunto(s)
Instrumentos Dentales , Preparación del Conducto Radicular , Temperatura , Falla de Equipo , Ensayo de Materiales , Aleaciones Dentales , Titanio , Diseño de Equipo , Estrés Mecánico